Evaporative Light-Scattering Detection, or ELSD for short, is a technology used with liquid chromatography to see UV-transparent (and UV-absorbing) compounds. In a previous post I talked about some applications where ELSD is not only useful, but required.
In this post, I will explain how an ELSD is configured and functions.
An ELS detector is quite different than most other detectors used in chromatographic applications. It does not require that compounds have any specific functionality or ionization capabilities. Its only requirements are that the compound does not evaporate under the conditions set on the detector and the solvent does. These conditions include setting evaporation temperature, inert gas flow, sample inlet flow, and sensitivity level.
So, what is the basis for the ELSD’s operation? Well, it is rather simple actually and can be explained in the following steps.
-The particles produced at the end of the drift tube are what are to be detected. so virtually any substance that will not evaporate in the drift tube and forms a particle or even an oil will be detected.
To help visualize this process I have included a graphic, which is representative of the Biotage® A-120 ELSD, Figure 1.
If you are interested in learning more about ELSD, please follow the link below to read more about the modular and fully automated ELSD solution, which seamlessly integrates with our Biotage® Selekt family of flash purification systems.