This application note describes a polymer-based strong anion exchange mixed-mode SPE protocol for the extraction of T3, rT3 and T4 from serum prior to LC-MS/MS analysis.


The method described in this application note demonstrates selective extraction of the thyroid hormones T3, rT3 and T4 from serum. 200 µL of serum was extracted using the EVOLUTE® EXPRESS AX 30 mg fixed well plate format. High reproducible recoveries and low phospholipid content were observed, demonstrating limits of quantitation < 50 pg/mL.